Agilent Technologies 1220 Infinity II LC User Manual page 68

Infinitylab lc series
Hide thumbs Also See for 1220 Infinity II LC:
Table of Contents

Advertisement

4
LAN Configuration
Dynamic Host Configuration Protocol (DHCP)
2 Set the Configuration Switch to DHCP either on the G1369C LAN Interface
Table 13
SW 4
ON
Table 14
SW 6
ON
3 Turn on the module that hosts the LAN interface.
4 Configure your Control Software (e.g. OpenLAB CDS ChemStation Edition,
68
Card or the main board of above mentioned modules.
G1369C LAN Interface Card (configuration switch on the card)
SW 5
SW 6
OFF
OFF
LC Modules with 8-bit configuration switch (B-firmware) (configuration switch
at rear of the instrument)
SW 7
SW 8
OFF
OFF
Lab Advisor, Firmware Update Tool) and use MAC address as host name,
e.g. 0030d3177321.
The LC system should become visible in the control software (see Note in
section
"General Information (DHCP)"
SW 7
SW 8
Initialization Mode
OFF
OFF
DHCP
Initialization Mode
DHCP
on page 66).
1220 Infinity II LC System User Manual

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents