Correcting Leaks
When
If a leakage has occurred in the flow cell area or at the heat exchanger or at the capillary
connections
Tools required
p/n
5043-0915
Preparations
Remove the front cover.
1 Use tissue to dry the leak sensor area and the leak pan.
2 Observe the capillary connections and the flow cell area for leaks and
3 Replace the front cover.
1220 Infinity II LC System User Manual
Description
Tissue
Wrench, 1/4 inch
for capillary connections
Fitting mounting tool
for bio-inert capillaries
correct, if required.
Figure 66
Observing for Leaks
Maintenance and Repair
Diode Array Detector (DAD)
11
321