Replacing Capillaries

Only perform this procedure if the HPLC-Chip Cube MS Interface (G4240A) is mounted
C A U T I O N
on the MS or sitting flat on a table as the center of gravity shifts and the instrument will
tip over without adequate support.
1 Press the cover release button once. When the lock opens
pull the front panel down.
HPLC-Chip Cube Interface User's Guide
When required
When broken, blocked or kinked
Tools required
Cube wrench, P/N G4240-83800
Hex key 3 mm, 12 cm long, P/N 8710-2411
Parts required
Fused silica/PEEK capillary (see Parts)
Preparations
• Unload the HPLC-Chip
Maintenance Procedures
2 Unlock the stages assembly and flip it out.
Replacing Capillaries