10
Error Information
DAD Detector Error Messages
takes longer and the detector uses special settings (e.g full data rate at 80 Hz
plus full spectra plus all signals) during data buffering.
Probable cause
1

Instrument Logbook

Figure 63
The logbook does not indicate a communication loss (power fail). It just shows the
N O T E
recovering (Power on, Lamps on).
240
CompactFlash Card is full.
Instrument Logbook
Suggested actions
• Correct communication problem.
• Reduce data rate.