Test Chromatogram

A built-in pre-defined test chromatogram on the VWM board is processed
through ADC like normal signals from the diodes and can be used to check the
ADC and the data handling system. The signal is available at the analog output
and on the GPIB.
The run time of the chromatogram is depending on the setting for response time (peak
N O T E
width). If no stop time is set the chromatogram will repeat continuously.
Response Time
0.06 sec
0.12 sec
0.25 sec
0.50 sec
1.00 sec
2.00 sec
4.00 sec
8.00 sec
The test chromatogram has four main peaks with the following absorbances:
Peak
1
2
3
4
1220 Infinity II LC System User Manual
Test Functions and Calibration
Variable Wavelength Detector (VWD)
Stop Time
0.8 min
0.8 min
0.8 min
0.8 min
1.6 min
3.2 min (Default settings)
6.4 min
12.8 min
Absorbance (approx.)
38 mAU
100 mAU
290 mAU
20 mAU
9