10
Error Information
Autosampler Error Messages
Autosampler Error Messages
These errors are specific to the autosampler.
Arm Movement Failed
Error ID: 4002
The transport assembly was unable to complete a movement in one of the
axes.
The processor defines a certain time window for the successful completion of
a movement in any particular axis. The movement and position of the
transport assembly is monitored by the encoders on the stepper motors. If the
processor does not receive the correct position information from the encoders
within the time window, the error message is generated.
See figure
• Arm Movement 0 Failed: X-axis.
Probable cause
1
2
3
216
Figure 35
on page 109 for axes identification.
Arm Movement 1 Failed: Z-axis.
Arm Movement 2 Failed: Theta (gripper rotation).
Arm Movement 3 Failed: Gripper (gripper fingers open/close).
Mechanical obstruction.
High friction in the transport assembly.
Defective motor assembly.
Suggested actions
Ensure unobstructed movement of the
transport assembly.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
1220 Infinity II LC System User Manual