Metering Home Failed

Error ID: 4054, 4704
The metering piston has failed to move back to the home position.
The home position sensor on the sampling unit flex board monitors the home
position of the piston. If the piston fails to move to the home position, or if the
sensor fails to recognize the piston position, the error message is generated.
Probable cause
1
2
3
4

Missing Vial

Error ID: 4019, 4034, 4541, 4706
No vial was found in the position defined in the method or sequence.
When the gripper arm picks a vial out of the sample tray, the processor
monitors the gripper motor encoder. If a vial is present, the closing of the
gripper fingers is limited by the vial. However, if no vial is present, the gripper
fingers close too far. This is sensed by the processor (encoder position),
causing the error message to be generated.
Probable cause
1
2
1220 Infinity II LC System User Manual
Defective sensor or main board.
Broken piston.
Defective metering-drive motor.
Defective main board.
No vial in the position defined in the
method or sequence.
Incorrect gripper alignment.
Error Information
Autosampler Error Messages
Suggested actions
Please contact your Agilent service
representative.
Exchange the metering piston and seal.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Install the sample vial in the correct position, or
edit the method or sequence accordingly.
Align gripper.
10
219