Missing Wash Vial; Motor Temperature - Agilent Technologies 1220 Infinity II LC User Manual

Infinitylab lc series
Hide thumbs Also See for 1220 Infinity II LC:
Table of Contents

Advertisement

10
Error Information
Autosampler Error Messages
Probable cause
3
4

Missing Wash Vial

Error ID: 4035, 4542, 4707
The wash vial programmed in the method was not found.
When the gripper arm picks a vial out of the sample tray, the processor
monitors the gripper motor encoder. If a vial is present, the closing of the
gripper fingers is limited by the vial. However, if no vial is present, the gripper
fingers close too far. This is sensed by the processor (encoder position),
causing the error message to be generated.
Probable cause
1

Motor Temperature

Error ID: 4027, 4040, 4261, 4451
One of the motors of the transport assembly has drawn excessive current,
causing the motor to become too hot. The processor has switched OFF the
motor to prevent damage to the motor.
See figure
• Motor 0 temperature: X-axis motor.
220
Defective gripper assembly (defective
gripper fingers or belt).
Defective transport assembly flex board.
No wash vial in the position defined in the
method.
Figure 35
on page 109 for motor identification.
Motor 1 temperature: Z-axis motor.
Motor 2 temperature: Theta (gripper rotation) motor.
Motor 3 temperature: Gripper motor (motor for gripper fingers).
Suggested actions
Exchange the gripper assembly.
Please contact your Agilent service
representative.
Suggested actions
Install the wash vial in the correct position, or
edit the method accordingly.