Error Information
Autosampler Error Messages
The autosampler attempted to initialize with a vial still in the gripper.
During initialization, the autosampler checks correct operation of the gripper
by closing and opening the gripper fingers while monitoring the motor
encoder. If a vial is still in the gripper when initialization is started, the
gripper fingers cannot close causing the error message to be generated.
Probable cause
1

Invalid Vial Position

Error ID: 4042
The vial position defined in the method or sequence does not exist.
The reflection sensors on the transport assembly flex board are used to
automatically check which sample trays are installed (coding on tray). If the
vial position does not exist in the current sample tray configuration, the error
message is generated.
Probable cause
1
2
3
218
Vial still in gripper.
Incorrect tray or trays installed.
Incorrect vial positions defined in the
method or sequence.
Tray recognition defective (dirty sample tray
or defective transport assembly flex board).
Suggested actions
Remove the vial using the Release Vial
function in the user interface. Reinitialize the
autosampler.
Suggested actions
Install the correct trays, or edit the method or
sequence accordingly.
Check method settings vs. installed tray.
Ensure the coding surfaces of the sample tray
are clean (located at the rear of the sample
tray). If this does not solve the problem,
contact your Agilent service representative.