Replacing Capillaries on a Standard Flow Cell

Replacing Capillaries on a Standard Flow Cell

For bio-inert modules use bio-inert parts only!
When
If the capillary is blocked
Tools required
p/n
OR
5043-0915
Parts required
Description
For parts see
Preparations
Turn the lamp(s) off.
Remove the front cover.
Remove the flow cell, see
All descriptions in this procedure are based on the default orientation of the cell (as it is
N O T E
manufactured). The heat exchanger/capillary and the cell body can be fixed mirror
symmetrically to have both capillaries routed to the bottom or to the top (depending on the
routing of the capillaries to the column).
The fittings at the flow cell body are special types for low dead volumes and not
N O T E
compatible with other fittings.
When retightening the fittings, make sure that they are carefully tightened (handtight plus
1/4 turn with a wrench). Otherwise damage of the flow cell body or blockage may result.
182
Description
Wrench, 1/4 inch
for capillary connections
Fitting mounting tool
for bio-inert capillaries
Wrench, 4 mm
(for capillary connections)
Screwdriver, Pozidriv #1 PT3
"Standard Flow Cell"
on page 208.
"Exchanging a Flow Cell"
on page 172.
Agilent 1260 Infinity DAD and MWD User Manual