Configuration Switches

The configuration switch can be accessed at the rear left side of the
instrument.
The instrument is shipped with switches 7 and 8 set to ON, which means that
the instrument is set to a default fixed IP address: 192.168.254.11
To configure the LAN, SW1 and SW2 must be set to OFF.
N O T E
Table 10
Initialization ('Init') Mode
Link Configuration
For the G4294B, the Configuration switches on the DAD main board must be used for
N O T E
configuring the LAN access of the instrument. The switches on the main board must all be
set to off.
1220 Infinity II LC System User Manual
Factory Default Settings
Using Default, switches 7 and 8 set to ON.
Speed and duplex mode determined by auto-negotiation
LAN Configuration

Configuration Switches

4
63