Replace a D2 Flow Chip (Bus Jumper)

Maintaining Your GC
1
Gather the following:
New gaskets
New D2 Flow chip (bus jumper)
2
Prepare for maintenance. On the GC touch display, go to
Maintenance > Instrument > Perform Maintenance > Maintenance
Mode> Start Maintenance.
3
When the GC has cooled and is ready to continue, open the
GC front door and lower the D2 front panel.
D2 front panel
GC front door
4
Remove the bus door and the D2 bus door.
Maintaining Columns and Bus Components
2
83