Configuration of the Instrument After an Upgrade Installation

Configuration of the Instrument After an Upgrade Installation

This step is only necessary if any of the following hardware upgrade kits has
been installed to the instrument.
• G4297A - 1220 Infinity Oven kit
• G4298A - 1220 Infinity upgrade manual injector to ALS
• G4299A - 1220 Infinity upgrade isocratic to gradient pump
1 Connect the Instrument to the Instrument Utility / Lab Advisor software.
2 Configure your Instrument according to the applied hardware changes:
3 Disconnect from the software and re-boot your instrument.
4 Re-connect to the Instrument to the Instrument Utility / Lab Advisor
5 Now perform the Installation Check as described in
6 When starting the Chromatographic Data System use Auto Configuration to
48
Software Revision B.01.04 and below:
Tools > Module Service Center (any module)
Software Revision B.02.01 and higher:
Instrument Control > Controls (any module) > Convert Device Type
(For example if you installed a G4297A - 1220 Infinity Oven kit then press
Add Oven
software.
Performing the 'Installation Check'"
have your new hardware configuration configured to the CDS.
"Priming the System and
on page 49.