Manual Injector
Overview of Maintenance Procedures
Table 45
Procedure
Flushing the injector
Exchanging the injection-valve
seal
Flushing the Manual Injector
The use of aqueous buffers or salt solutions can lead to crystal formation.
C AU T I O N
Crystal formation may cause scratches on the injection seal.
➔
1 Switch the valve to the INJECT position.
2 Use the pump to flush the sample loop and seal grooves.
3 Use the needle-port cleaner (supplied with the valve) and syringe to flush
1220 Infinity II LC System User Manual
Overview of maintenance procedures
Always rinse the valve with water after using aqueous buffers or salt solutions.
the needle port and vent capillary.
Typical frequency
After using aqueous buffers or
salt solutions
After approximately 10000 to
20000 injections, or when the
valve performance shows
indication of leakage or wear
Maintenance and Repair
Manual Injector
Time required
5 min
10 min