9
Place the pump housing onto the piston housing.
11
Tighten the lock screw.
1220 Infinity II LC System User Manual
10
Insert the pistons and carefully press them into the
seals.
12
Install the pump head, see
Assembly"
on page 262.
Maintenance and Repair
Solvent Delivery System
"Reinstalling the Pump Head
11
261