Replace the fan cover (1.) and fix its screw (2.).
Next Steps:
11
Reset the lamp counter as described in the Utilities software documentation.
12
Turn the lamp ON.
13
Give the lamp more than 10 min to warm-up.
14
Perform Wavelength Calibration to check the correct positioning of the lamp.
1220 Infinity II LC System User Manual
2.
1.
Variable Wavelength Detector (VWD)
10
Close the lower front cover.
Maintenance and Repair