Install the Restriction Capillary

When
Tools required
p/n
8710-0510
Parts required
Description
Restriction capillary
Preparations
Remove the column, see
1
Open both front covers.
3
Connect the restriction capillary to the injector valve
(port 6).
1220 Infinity II LC System User Manual
Description
Wrench open 1/4 — 5/16 inch
"Remove a Column"
2
Place the restriction capillary inside the column
compartment.
4
Connect the restriction capillary to the flow cell inlet of
the detector.

Install the Restriction Capillary

on page 55.
3
Installation
57