3
Installation
Installing the Hardware
6 Place the Solvent Inlet Filter end of the Bottle Head Assembly in the Solvent
7 Prime the tubing using the Syringe (9301-044) and Syringe adapter
8 Connect the Bottle Head Assembly connector (see
9 Connect the waste tube with the fitting attached (part of Accessory Kit) to
10 Attach the corrugated waste tube (part of Accessory Kit) to the VWD leak
11 Connect the waste tube (part of Accessory Kit) to the purge valve outlet
12 Connect the network connection between the instrument and your PC.
More details about how to establish a network connection to the instrument can be found
N O T E
in
Configuration"
13 Verify that the power push button at the front of the module (see
40
Bottle (see picture below).
Figure 4
Bottle Head Assembly and Solvent Bottle
(9301-1337) (part of the Accessory kit) until the tubing is completely filled
with water.
1+2) to:
the passive inlet valve (isocratic pump) or
the degasser inlet Channel A (gradient pump).
the flow cell outlet and the other end to an appropriate solvent waste
container (see
Figure 5
tray outlet adapter and guide it to a proper waste container (see
page 41).
adapter and the other end to the waste container.
"Connecting the Agilent 1220 Infinity II LC to the PC"
on page 59
on page 41) stands off. Now connect the power cord to the instrument and
the power line.
on page 41).