Exchanging the Needle Seat Assembly

When
When the seat is visibly damaged
When the seat capillary is blocked
Tools required
Description
Wrench, 1/4 inch (supplied in HPLC Tool-Kit)
Screwdriver, flat-head
Parts required
#
1
Preparations
• Select Start in the Tools function in LMD Software Change Needle
• Remove the upper front cover.
• Use the Needle Up command in the Change Needle function to lift the needle an additional
1
Disconnect the seat-capillary fitting from the injection
valve (port 5).
278
p/n
Description
G1313-87101
Needle-seat assy (0.17 mm i.d. 2.3 µL)
1 cm.
2
Use a small flat-head screwdriver to ease out the needle
seat.
1220 Infinity II LC System User Manual