9
Test Functions and Calibration
1220 Infinity II LC System
Module Info
The Module Info tool collects diagnostic information from a module and writes
the results to a file. You can view the results in three tabs:
General
The General tab shows information about the module's firmware and options in
a two-column table.
Tables
The Tables tab allows you to display the contents of all available diagnostic
tables for the module. You click the [+] sign to open a table, or the [-] sign to
close an open table.
Signals
The Signals tab shows the plots of the available diagnostic signals from the
module. The signal plots that are available are module-dependent; where
available, both short-term and long-term plots are displayed for a signal.
Diagnostic Buffers
Each module contains a set of diagnostic buffers that can be queried.
Lab Advisor allows you to query all diagnostic buffers for a selected module
and write the results to a text file. The text file is written in English.
State Info
The State Info tool displays the current status of all Agilent 1220 Infinity II LC
modules; the status is continuously updated. Unless aborted, the tool runs
continuously for 60 min. Click Stop Test to stop the tool.
144
1220 Infinity II LC System User Manual