Remove and Install a Flow Cell

When
If an application needs a different type of flow cell or the flow cell needs repair.
Tools required
Description
Wrench, 1/4 inch
for capillary connections
Parts required
p/n
G1315-60022
G1315-60025
G1315-60024
G1315-60015
Preparations
Turn the lamp(s) off.
1
Remove the lower front cover.
1220 Infinity II LC System User Manual
Description
Standard flow cell, 10 mm, 13 µL, 120 bar (12 MPa)
Semi-micro flow cell, 6 mm, 5 µL, 120 bar (12 MPa)
Micro flow cell, 3 mm, 2 µL, 120 bar (12 MPa)
High pressure flow cell, 6 mm, 1.7 µL, 400 bar (40 MPa)
Nano flow cell
2
Open the flow cell door.
Maintenance and Repair