3 Record the hardware (MAC) address (for example, 0010835675AC).
4 The Error means the MAC address has not been assigned an IP address and
5 Close the log file before turning on another instrument.
6 Uncheck the Do you want to log BootP requests? box after configuring
Adding each instrument to the network using BootP
1 Follow Start > All Programs > Agilent BootP Service and select Edit BootP
2 Uncheck the Do you want to log BootP requests? once all instruments have been
3 Click Edit BootP Addresses... The Edit BootP Addresses screen appears.
4 Click Add... The Add BootP Entry screen appears.
1220 Infinity II LC System User Manual
the Tab File does not have this entry. The MAC address is saved to the Tab
File when an IP address is assigned.
instruments to avoid having the logfile use up excessive disk space.
Settings. The BootP Settings screen appears.
added.
The Do you want to log BootP requests? box must be unchecked when you have
finished configuring instruments; otherwise, the log file will quickly fill up
disk space.
Figure 17
Enable BootP logging
LAN Configuration
Automatic Configuration with BootP
4
75