Maintenance of Standard, Semi-Micro or Micro Flow Cell

Maintenance of Standard, Semi-Micro or Micro Flow Cell

For bio-inert modules use bio-inert parts only!
When
If the flow cell needs repair due to leaks or contaminations (reduced light throughput)
Tools required
p/n
OR
5043-0915
Parts required
Description
For parts, see
"Micro Flow Cell"
Preparations
Turn the flow off.
Remove the front cover.
Remove the flow cell, see
The gaskets used in the standard and semi-micro/micro flow cell are different.
N O T E
Sample degradation and contamination of the instrument
C AU T I O N
Metal parts in the flow path can interact with the bio-molecules in the sample leading
to sample degradation and contamination.
176
Description
Wrench, 1/4 inch
for capillary connections
Fitting mounting tool
for bio-inert capillaries
Hexagonal key, 4 mm (supplied in HPLC Tool-Kit)
Toothpick
"Standard Flow Cell"
on page 214.
"Exchanging a Flow Cell"
For bio-inert applications, always use dedicated bio-inert parts, which can be
identified by the bio-inert symbol or other markers described in this manual.
Do not mix bio-inert and non-inert modules or parts in a bio-inert system.
on page 208,
"Semi-Micro Flow Cell Parts"
on page 172.