Exchanging A Flow Cell - Agilent Technologies 1100 Series Reference Manual

Variable wavelength detector
Hide thumbs Also See for 1100 Series:
Table of Contents

Advertisement

4
Repairing the Variable Wavelength Detector

Exchanging a Flow Cell

When required
Tools required
Parts required
Preparations for this procedure:
Turn the lamp off.
88 1100 Series Variable Wavelength Detector Reference Manual
If application needs a different type of flow cell or the flow cell needs
repair.
Two 1/4 inch wrenches for capillary connections
Standard Flow Cell G1314-60086 10 mm, 14 µl, 40 bar,
( former Standard Flow Cell G1314-60080 is obsolete )
Micro flow cell, 5 mm, 1 µl, 40 bar, G1314-60081
Semimicro flow cell, 6 mm, 5 µl, 40 bar, G1314-60083
High pressure flow cell, 10 mm, 14 µl, 400 bar, G1314-60082