Pressure Cluster Partner Missing; Position Cluster Partner Missing - Agilent Technologies 1290 Infinity User Manual

Valve drive and valve heads
Hide thumbs Also See for 1290 Infinity:
Table of Contents

Advertisement

6
Error Information
Module Specific Error Messages

Pressure Cluster Partner Missing

The connection from the valve drive to a defined pressure cluster partner is
lost.
Probable cause
1
2

Position Cluster Partner Missing

Probable cause
1
2
74
Communication issues
Configuration mismatch
Communication issues
Configuration mismatch
Suggested actions
Check the CAN cable connections of the
modules.
Check and correct if necessary the valve
configuration and presence of defined pressure
cluster partner.
Suggested actions
Check the CAN cable connections of the
modules.
Check and correct if necessary the valve
configuration and presence of defined position
cluster partner.
Agilent 1290 Infinity Valve Drive and Valve Heads

Advertisement

Table of Contents
loading

This manual is also suitable for:

G1170a

Table of Contents