Needle Carrier Failed; Missing Vial Or Missing Well-Plate - Agilent Technologies 1260Infinity II Bio-inert FractionCollector User Manual

Infinitylab lc series
Table of Contents

Advertisement

6
Error Information
Module Error Messages

Needle Carrier Failed

Error ID:
The needle carrier on the transport unit assembly failed to move correctly.
Probable cause
1
2
3
4
5

Missing Vial or Missing Well-plate

Error ID:
No vial or well-plate was found in the position defined in the method or
sequence.
When the needle carrier moves to a vial or well-plate and the needle is lowered
into the vial or well-plate, the position of the needle is monitored by an
encoder behind the vial pusher. If no vial or well-plate is present, the encoder
detects an error and the message "missing vial or well plate" is generated.
Probable cause
1
2
3
4
104
Defective position sensor in the needle
carrier assembly.
Bad needle carrier positioning in X or Theta.
Disconnected needle carrier connector.
Defective MTP main board.
Defective Z-motor.
No vial in the position defined in the
method or sequence.
Defective needle carrier assembly.
Defective transport unit assembly flex
board.
Defective MTP main board.
1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual
Suggested actions
Exchange the needle carrier assembly.
Perform a self-alignment.
Connect needle carrier connector correctly.
Please contact your Agilent service representative.
Please contact your Agilent service representative.
Suggested actions
Install the sample vial in the correct position, or
edit the method or sequence accordingly.
Exchange the needle carrier assembly.
Please contact your Agilent service representative.
Please contact your Agilent service representative.