7
Error Information
Pump Error Messages

Seal wash pump was missing when tried to turn on

Error ID: 29223
The seal wash pump has not been detected (while being configured or detected
before)
Probable cause
1
2
3

Valve hardware overcurrent (MCGV)

Error ID: 29227
Power consumption too high for one of the MCGV valves.
Probable cause
1
2
3
1260 Infinity II Flexible Pump User Manual
Defect cable connection to seal wash pump. Check cable connection.
Defect seal wash pump motor.
Defective main board.
Cable defect.
Valve defect
Defective mainboard.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Suggested actions
Replace MCGV/4CGV.
Replace MCGV/4CGV.
Please contact your Agilent service
representative.