Missing Vial

Error ID: 4019, 4034, 4541, 4706
No vial was found in the position defined in the method or sequence.
When the gripper arm picks a vial out of the sample tray, the processor
monitors the gripper motor encoder. If a vial is present, the closing of the
gripper fingers is limited by the vial. However, if no vial is present, the gripper
fingers close too far. This is sensed by the processor (encoder position),
causing the error message to be generated.
Probable cause
1
2
3
4
Agilent 1260 Infinity Standard Autosampler User Manual
No vial in the position defined in the method
or sequence.
Incorrect gripper alignment.
Defective gripper assembly (defective
gripper fingers or belt).
Defective transport assembly flex board.
Error Information
Autosampler Error Messages
Suggested actions
Install the sample vial in the correct position, or
edit the method or sequence accordingly.
Align gripper.
Exchange the gripper assembly.
Please contact your Agilent service
representative.