Usb Communications - Keithley 2461 Reference Manual

1kw pulse mode interactive sourcemeter instrument
Table of Contents

Advertisement

Model 2461 Interactive SourceMeter® Instrument Reference Manual
You can also try restarting the computer and the instrument. To restart the instrument:
1. Turn the instrument's power off, and then on.
2. Wait at least 60 seconds for the network configuration to be completed.
3. Press the MENU key.
4. Under System, select Communication.
5. Select the LAN tab.
6. Verify the settings.
If the above actions do not correct the problem, contact your system administrator.

USB communications

To use the rear-panel USB port, you must have the Virtual Instrument Software Architecture (VISA)
layer on the host computer. See
information.
VISA contains a USB-class driver for the USB Test and Measurement Class (USBTMC) protocol that,
once installed, allows the Microsoft
When you connect a USB device that implements the USBTMC or USBTMC-USB488 protocol to the
computer, the VISA driver automatically detects the device. Note that the VISA driver only
automatically recognizes USBTMC and USBTMC-USB488 devices. It does not recognize other USB
devices, such as printers, scanners, and storage devices.
In this section, "USB instruments" refers to devices that implement the USBTMC or
USBTMC-USB488 protocol.
The full version of National Instruments (NI
other kind of USB device that you want to communicate with VISA. For more information, see the
National Instruments
Using USB
A USB cable is shipped with the instrument. If the original cable is not available, you will need a USB
cable with a USB Type B connector end and a USB type A connector end. You will need a separate
USB cable for each instrument you plan to connect to the computer at the same time using the USB
interface.
To use a USB connection:
1. Connect the Type A end of the cable to the host computer.
2. Connect the Type B end of the cable to the instrument.
3. Turn power to the instrument on.
4. When the host computer detects the new USB connection, the Found New Hardware Wizard
starts.
5. On the "Can Windows connect to Windows Update to search for software?" dialog box, click No,
and then click Next.
6. On the "USB Test and Measurement device" dialog box, click Next, and then click Finish.
2461-901-01 A/November 2015
How to install the Keithley I/O Layer
®
Windows
®
) VISA provides a utility to create a USB driver for any
(http://www.ni.com) website.
(on page 2-79) for more
®
operating system to recognize the instrument.
Section 2: General operation
2-69

Advertisement

Table of Contents
loading

Table of Contents