Shields And Guarding; Figure 2-11 Basic Connections To Dut - Keithley 6220 Reference Manual

Ac and dc current source measure unit
Table of Contents

Advertisement

2-16
Output Connections
Model 6220/6221 Reference Manual
Figure 2-11
Basic connections to DUT
A) Inner shield connected to Output Low
B) Inner shield connected to Cable Guard

Shields and guarding

Noise shield connections
Figure 2-12
shows typical noise shielding for the two triax connector
configurations. A noise shield is used to prevent unwanted signals from being
induced into the test circuit. A test current below 1μA may benefit from effective
shielding. Typically, the metal noise shield surrounds the DUT and is connected to
Output Low.
Safety shield connections
A safety shield must be used whenever hazardous voltages (>30Vrms, 42Vpeak)
will be present. The metal safety shield must completely surround DUT test
circuitry and must be connected to a known Safety Earth Ground and Chassis
(see
Figure
2-13). Use #18AWG wire or larger for connection to safety Earth
Ground and Chassis.