Period Hours
Thermo Fisher Scientific
Next Time
The Next Time screen is used to view and set the next zero/span check date
and time. Once the initial zero/span check is performed, the date and time
of the next zero/span check is calculated and displayed.
In the Main Menu, choose Calibration > Zero/Span Check > Next
●
Time.
NEXT DATE AND TIME:
19 MAR 2005
12:34
PRESS
TO EDIT
RANGE
AVG
DIAGS ALARM
The Zero/Span Period Hours screen defines the period or interval between
zero/span checks. Periods between 0 and 999 hours are acceptable. To turn
the zero/span check off, set the period to 0.
In the Main Menu, choose Calibration > Zero/Span Check > Period
●
Hours.
ZERO/SPAN CHECK:
>NEXT TIME 01Jan05 12:00
PERIOD HR
TOTAL DURATION HR
ZERO DURATION MIN
SPAN DURATION MIN
PURGE DURATION MIN
ZERO/SPAN AVG SEC
RANGE
AVG
DIAGS ALARM
ZERO CAL RESET
SPAN CAL RESET
ZERO/SPAN RATIO
24
2.5
30
30
30
60
ON
OFF
1
NEXT DATE AND TIME:
19 MAR 2005
SETTING: DAYS
SET MONTHS
CHANGE VALUE
SAVE VALUE
RANGE
AVG
DIAGS ALARM
Model 48i Trace Level-Enhanced Instruction Manual
Operation
Calibration Menu