Calibration Of The Falling Ball Viscometer - Thermo Scientific HAAKE Instruction Manual

Table of Contents

Advertisement

periodically:
with nationally recyclable
standards:
Order-No.
082--5042
082--5043
082--5044
082--5046
082--5336
Order-No.
082--5303
082--5304
082--5305
Calibration

11. Calibration of the Falling Ball Viscometer

The falling ball viscometer is checked according to the
HAAKE regulations before delivery. The geometry relates
strictly on the DIN 53015/ISO 12058 and therefore the listed
ball constants can be used.
The calibration of the instrument can be done using nation-
ally recyclable calibration fluids. This procedure is also rec-
ommended for new balls or after repairs and for falling ball
viscometers integrated into a measuring procedure accord-
ing to ISO 9000.
Testing ...
to be defined by the user or to be read in the QC-hand book.
Calibration fluids acknowledged in the corresponding user
country, e.g. Europe: DKD/PTB calibration fluids, in Japan:
JIS fluids or in the USA: Cannon calibration fluids.
The calibration excludes the influence of allowed tolerances
when measuring balls and falling tubes.
The following Thermo Fisher Scientific products can be used
for testing the instrument:
Type
E7
E200
E2000
E6000
E15000
For calibration of the falling ball viscometer the following cal-
ibration fluids (recyclable in Europe) are recommended:
Type
100 BW
2000 AW
10000 B
18
 (mPas)
for balls
at 20C
5
120
1900
6000
15000
 (mPas)
for balls
at 20C
100
2000
10000
1
2, 3
4
5
6
2, 3
4, 5
5, 6