Restore Default Flow Calibration; Input Board Calibration - Thermo Scientific 42i Instruction Manual

Chemiluminescence no-no2-nox analyzer
Hide thumbs Also See for 42i:
Table of Contents

Advertisement

Operation
Service Menu
Restore Default Flow

Input Board Calibration

3-80
Model 42i Instruction Manual
Note Wait at least 30 seconds for the reading to stabilize before saving the
value.
In the Main Menu, choose Service > Flow Calibration > Span.
The Restore Default Flow Calibration screen allows the user to reset the
flow calibration configuration values to factory defaults.
Calibration
In the Main Menu, choose Service > Flow Calibration > Set Defaults.
RESTORE DEFAULT CAL:
RANGE
AVG
DIAGS ALARM
The Input Board Calibration menu is used to initiate a calibration of the
input board A/D stages. The input board calibration menu is visible only
when the instrument is in service mode. For more information on the
service mode, see
Note This adjustment should only be performed by an instrument service
technician.
In the Main Menu, choose Service > Input Board Calibration.
CALIBRATE FLOW SPAN:
CURRENTLY:
SET TO:

MOVE CURSOR

CHANGE VALUE
RANGE
AVG
DIAGS ALARM
RESTORE
"Service
Mode" earlier in the chapter.
1.000 LPM
0.800
LPM ?
SAVE
RESTORE DEFAULT CAL:
ARE YOU SURE YOU WANT
PRESS
TO CONFIRM RESTORE
RANGE
AVG
RESTORE
TO?x
DIAGS ALARM