Setting The Internal Standard Correction - Agilent Technologies 7500 ICP-MS Operator's Manual

Table of Contents

Advertisement

N O T E
N O T E
1 Select an interpolation formula in the Mass Interpolation Fit: drop-down list in
N O T E
Agilent 7500 ICP-MS ChemStation Operator's Manual
Performing a Quantitative Analysis
To save edited calibration information, the calibration must be saved.
For online data analyses, select Top window >> Method >> Save Calibration. For
offline data analyses, select Offline Data Analysis window >> Method >> Save
Calibration.

Setting the Internal Standard Correction

ChemStation allows you to set an internal standard for each element.
Select the internal standard closet to the mass of the analyte. The element set as an
internal standard will not be quantitatively analyzed.
The CPS of an analyte can be corrected based on the value (VIS) calculated using an
interpolation formula without changing the existing internal standard correction.
The internal standard correction can be set using the following procedure.
If an internal standard is not added to the standard sample and sample, the following
step is not necessary.
the upper right section of the Edit Calibration dialog box.
Refer to "Interpolation Formulas for Virtual-Internal-Standard Correction" in
"Quantitation" in Appendix B "Formulas." for details on interpolation formulas.
If VIS is not used, it is not necessary to select an interpolation formula.