Agilent 7500 ICP-MS ChemStation Operator's Manual
Configuration
• Remote Shutdown
Standard (Default)
The Standard option must be selected when the ICP-MS is connected to the ISIS.
Agilent LC
The Agilent LC option must be selected when ISIS is used with an external
pump.
The Agilent LC option can be selected when the ICP-MS is connected to any
modular Agilent LC (eg. Agilent 1100, Agilent 1050) using an APG remote
cable. This option will enable bidirectional communication between the LC and
the ICP-MS in the event of a hardware error on either the LC or the ICP-MS. If
there is a hardware error on the ICP-MS, or if the plasma is turned off via
sequencing, a shutdown command will be sent to the LC. Also, if there is any
hardware error on the LC, a shutdown command will be sent to the ICP-MS,
which will turn off the ICP-MS plasma and switch the instrument from Analysis
mode to Standby mode.
• Maintenance Log
Record Log Every Tuning Report Generation
When the option "Record Log Every Tuning Report Generation" is selected,
printing a tuning report or selecting a tuning report in Autotune will cause the
tuning parameters (as well as all parameters associated with meter indications
and the status of devices) to be recorded in a maintenance log.
• Processing during ALS probe movement
Stop PeriPump rotation
The peristaltic pump will stop when the auto sampler probe is in motion, this will
prevent air from being taken in from the sample probe.
Change Carrier Gas Flow Rate
Flow rate: Initially 0.2 l/min; set in a range from 0.0 to 2.00 l/min Stabilization
time: Initially 12 sec; set in a range from 0 to 999 sec
While the auto sampler probe is in motion, the carrier gas flow rate is changed to
the set value. If a stabilization time is set, the probe is moved to the specified
cleaning port after expiration of the set time.