Agilent 7500 ICP-MS ChemStation Operator's Manual
Creating a Method
1 Add, edit, or delete an element.
• When adding or editing, right-click on the element in Element List Information,
and choose an option from the drop-down menu. Set the mass number (only when
adding), the element name, and the integration time. In the Multi-Tune mode, set
the tune step as well. The detector mode is set to "Auto" when an element is
added, and is not changed when an element is edited.
Adding an Element
• When deleting, select an element in Element List Information and right-click on
it. Choose Delete on the menu.
2 View or edit an interference correction equation.
• When viewing, select the interference correction equation in Interference
Equations and click on Edit/View Equation.
• When selecting another equation, select the interference correction equation in
Interference Equations.
• When editing an equation, select the interference correction equation in
Interference Equations and click Edit/View Equation. For the procedure for
editing interference correction equations, see "Creating and Editing Interference
Correction Equations" in "Using the Default Method as a Template" below. The
change will be reflected in the list box.
Interference-Equation Dialog Box
5-8
Editing an Element