Specifying An Element For Report Output; Saving Parameters - Agilent Technologies 7500 ICP-MS Operator's Manual

Table of Contents

Advertisement

Agilent 7500 ICP-MS ChemStation Operator's Manual
Performing a Semiquantitative Analysis

Specifying an Element for Report Output

In semiquantitative analysis, it is possible to control the inclusion of individual
elements in the report. For instance, elements receiving interference from oxides,
doubly charged ions, etc can be excluded from the report.
Output settings can be entered using the following procedure.
1 In the list displayed in the Report area, select the elements that you do not want
to include in the report.
In the default setting, all elements are shown in the list.
Select multiple consecutive elements by holding down the Shift key.
Select any multiple elements by holding down the Ctrl key.
2 Click Delete Selected Masses.
The selected elements will be deleted from the list.
3 Select the elements/isotopes to add to the report from the drop-down list and
click Add.
The selected elements will be added to the list.
4 To load elements from a method, click Load Masses from Method.
5 To load masses from the correction list on the left side, click Load Masses from
Correction List. All masses selected in the correction list will be loaded to the
report list.
Elements will be loaded from the method.

Saving Parameters

Set semiquantitative parameters can be saved using the following method.
1 Click Save Parameters.
The Save SemiQuant Parameter File dialog box will appear.
Save SemiQuant Parameter File Dialog Box
13-15