1 Select File>>Load Tune Values.
2 Select the tune file in one of the following ways:
N O T E
Agilent 7500 ICP-MS ChemStation Operator's Manual
Tuning

Loading a Tune File

To use a saved tune file for a tuning procedure, load the file by completing the
following steps in the Tune window:
The Load Tuning Parameter File dialog box appears.
Load Tuning Parameter File Dialog Box
• Type the name of the tune file and click OK.
• Double-click a tune file in the displayed list.
• Click a tune file in the displayed list and click OK.
Agilent 7500 ChemStation will immediately change tuning parameters according to
the selected tune file.
If you change parameter values during the tuning procedure and want to save the new
values, save the tune file again. Otherwise, the changes are saved in atune.u only.
When you save, you can either overwrite or make a new tune file by renaming it.