Tuning Solution; Tuning Procedure - Agilent Technologies 7500 ICP-MS Operator's Manual

Table of Contents

Advertisement

N O T E
1 Install the Shield plate
2 Select Tune>>Acq. Params on the Tuning window.
3 Aspirate a tuning solution (1 ppb of Mg, Y and Tl).
4 Set the tuning parameters as typical values.
Agilent 7500 ICP-MS ChemStation Operator's Manual
Tuning

Tuning solution

Mg, Y, Ce, and Tl, solution at 1 ppb
For the high sensitivity mode, Li is not used because it is difficult to reduce the
background. Li remains on the interface as memory.

Tuning procedure

To tune the sensitivity, complete the following steps:
The Acquisition Parameters - Sensitivity dialog box appears. Change masses to 24,
89 and 205 and click OK.
Set the parameters as follows: