14 The message asking for the Agilent 7500 to be placed in Shutdown mode is
N O T E
15 Reboot the computer.
16 Install the Patch File if necessary. Refer to page 18-35.
17 Select Start>>All Programs>>ICP-MS ChemStation>>Edit hosts. Enter the IP
18 Reboot the computer.
19 Turn off the power of the Agilent 7500. Confirm the cable connection between
C A U T I O N
Agilent 7500 ICP-MS ChemStation Operator's Manual
Installing the Agilent 7500 ChemStation Software (Windows XP)
displayed. Click OK.
If the instrument is not in shutdown mode, press the shutdown switch on the
underside of the right instrument cover. The instrument transfers to shutdown mode
(it takes approximately 5 minutes). When it is in shutdown mode, the status LED
indicator on the cover turns off.
Address and the Host Name of Agilent 7500 which is setup in the Bootp Service.
Save the hosts file.
the Agilent 7500 and the computer. Then turn on the Agilent 7500.
After restarting the computer, turn off and turn on the Agilent 7500.
18-32