Setting Up an Acquisition Method

Setting Up an Acquisition Method

1 Set up the method in the Method Editor window:
a Enter or select the values and settings for each of the tabs below.
b Optional. To download the settings to the instrument, click
c To save the method, click
d Name the method and click OK.
2 Enter values for all of the LC modules configured for the instrument.
3 Enter the TQ parameter values.
a On the Acquisition tab, select the Scan type from the list in the Time
Agilent Triple Quadrupole LC/MS System User Guide
Segments table. The Acquisition Parameters table is cleared when you
change the Scan type. The parameters available on the right change
depending on the Scan type for the selected Time Segment.
If you are changing the Scan type from MRM to Dynamic MRM or to
Triggered MRM, you can copy and paste the transitions from the original
Scan segments table to the Clipboard and then to the new Scan segments
table. See the online help for more information.
Save As.
Apply.