Agilent Technologies 7500 ICP-MS Operator's Manual page 307

Table of Contents

Advertisement

Agilent 7500 ICP-MS ChemStation Operator's Manual
Setting Up a Sequence
2) To select a concentration from the calibration levels, choose the Select a
Calibration Level radio button and select a level from the drop-down list.
Proceed to step 5).
3) To set a specific value for the current sample, select the Set Specific Value radio
button.
4) Set a concentration in the Conc field of the Current ISTD Conc table.
The Current ISTD Conc table shows the internal standard currently set for
calibration. Click on the Conc field and enter a concentration value.
Click the right mouse button to display the popup menu, which allows you to copy,
paste, or delete the field.
Click Browse to set the concentration setting for each calibration level in the
Concentration field.
5) Click OK.
The Set ISTD Conc dialog box will close and the ISTD concentration level will be
set in the Edit Sample Log Table dialog box.
If the Select a Calibration Level radio button was selected, the level number will be
displayed. If the Set Specific Value radio button was selected, "Manual" will be
displayed.
9 To execute an action other than the action set as the method in case of a QC
check failure, select a process to be executed from the Action on Failure
drop-down list.
10 To skip a method in the sequence, double-click on Skip and select Skip.
Double-clicking Skip will alternately display and hide Skip.
11 To use the LC/GC ChemStation, set a method and vial number to use with the
LC/GC ChemStation.
1) Click the right mouse button and select Show Columns for LC/GC ChemStation
in the popup menu.
2) Double-click the LC/GC Method field, then select a desired LC/GC ChemStation
method from the Select File dialog box displayed.
If you leave the LC/GC Method field blank, the method already loaded in the
LC/GC ChemStation will be used during the sequence.
6-9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents