Agilent 7500 ICP-MS ChemStation Operator's Manual
Creating a Method

Editing Element Information

Element List Information on the Elements page shows the mass numbers and
names of the elements for which data is to be acquired. When the acquisition mode
is set to "Multi Tune," the tune mode also appears.
In the Multi Tune mode, tune files with the same names as the tune names shown on
this page shall be prepared before the method is actually used. The tune names
represent the modes specified below. As the modes using a reaction gas, "HE" is
helium-gas mode, "H2" is hydrogen-gas mode, "H2MOD" is
modified-hydrogen-gas mode. "NOGAS" and "NORM" are modes that do not use a
reaction gas. "COOL" is cool plasma and "HOT" is hot plasma. Depending on the
instrument type, application, and sample, only the appropriate tune names are
shown. For details on tuning, see Chapter 4, "Tuning."
Interference Equations show the library names of all interference correction
equations. An explanation of the selected item appears at the bottom of the screen.
Elements Page
Edit the element information by the following procedure.
5-7