Agilent 7500 ICP-MS ChemStation Operator's Manual
Configuration
7 Check EM Protection
Normally, leave Auto setting of integ time in analog mode selected. If the detector
mode is set to auto, the dwell time in analog mode is fixed at 100 microseconds so
that large signals do not damage the detector, regardless of the dwell time value
specified in the method.
In Spectrum and Isotope analysis, output counts are calculated and displayed as cps.
In Time Resolved analysis, output counts are calculated to get counts for the
integration time set in the acquisition parameters of the method. When Auto setting
of integ time in analog mode isn't selected and the detector mode is set to AUTO in
the acquisition parameters, the acquisition will be executed according to the
integration time; the integration time doesn't change
8 Select a QC mode
Click Browse, and select a QC mode file. QC mode files have qcc extensions.
User-defined QC modes can be created by selecting Advanced QC.
For more information about the QC configuration, refer to the Intelligent Sequence
manual.
Open File Dialog Box
2-7