Agilent 7500 ICP-MS ChemStation Operator's Manual
Creating a Method
Interference Information provides information about isobaric interferences where
elements have isotopes that overlap at the same mass. You can also view possible
interferences from polyatomic ions by clicking in the check boxes at the bottom of
the dialog box. Polyatomic ions of the selected specific type that could interfere with
the masses of the element will be displayed. Multiple types of polyatomic ions can
be selected.
ICP-MS DATA BASE (Interference Information) Dialog Box
N O T E
Click Prev! to see interference information about the previous element (in this case
Ga). Click Next! to see interference information about the next element (in this case
As). Click Index to see information about any other elements. An element index will
pop up. Click on the element then click OK.