Agilent 7500 ICP-MS ChemStation Operator's Manual
Startup, Shutdown and Status
(continued)
Meter
Shut Down Mode
Carrier Gas Flow
Makeup Gas Flow
Optional Gas Flow
Plasma Gas Flow
Aux Gas Flow
Forward Power
Reflected Power
*RP: Rot
ary Pump
**BB: Babington Nebulizer, CF: Cross-flow Nebulizer, CN: Concentric Nebuli
Ion Lenses Test
This test will check to determine if any of the ion lenses are shorted.
1 Verify that the ion lenses are connected to the equipment.
If the ion lenses are not connected, the test results will not be valid because the ion
lenses will not be detected as being shorted.
2 Place the equipment in standby mode and verify that the cover switch is closed.
3 Select Instrument>> Instrument Control from the the Top menu.
The ICP-MS Instrument Control window will be displayed.
4 Select Diagnostics >> Ion Lenses Test.
Typical Range
Standby Mode
0
0
0
0
0
0
0
0
0
0
0
0
0
0
Unit
Analysis Mode
0.8 to 1.3
L/min
0 to 1.0
L/min
Depends on
application
15
L/min
0 to 1.0
L/min
700 to 1600
< 20
zer
%
W
W
3-19