Agilent Technologies 7500 ICP-MS Operator's Manual page 207

Table of Contents

Advertisement

3 You can select the data file output format.
4 Click the Data Acquisition check box.
5 Click the Data Analysis check box.
N O T E
6 Click OK to enter the changes.
Agilent 7500 ICP-MS ChemStation Operator's Manual
Creating a Method
• Export AIA format for Agilent LC/GC checkbox
Selecting this checkbox will cause the data file to be saved in AIA format as well.
This file is saved to the data file folder. Use the Agilent LC/GC ChemStation to
import the file for display or analysis.
• Export Agilent LC/MSD raw data checkbox
Selecting this checkbox will cause the data file to be saved as LC/MSD raw data
as well. This file is saved to the data file folder. Use the Agilent LC/MSD
ChemStation to load, display, or analyze the file.
The default method should show this check box already marked. For data acquisi-
tion to occur, you must click the check box if it is not already marked.
Clicking this box tells ChemStation to run the data analysis section of the method
following data acquisition.
The pre-and post-run macro command check boxes enable you to execute actions
before or after acquisition. Macros included with ChemStation allow automatic
instrument startup and shutdown. Contact Agilent Technologies engineers for
information about these as well as additional customized macros.
The Interference Equation dialog box appears. To continue creating the equation,
see "Creating/Editing Interference Correction Equations" in the next section.