1 Confirm that the currently loaded method at the Top level is saved in the same
N O T E
2 Select Sequence
Agilent 7500 ICP-MS ChemStation Operator's Manual
Running a Sample Analysis

Running a Sequence

Running a sequence starts the sample analysis. To run a sequence that is already
loaded, complete the following steps from the Top window:
directory as the method to be used for reprocessing.
If not, load the method to be used for reprocessing by using Methods
Confirm that the edited method and calibrations is saved.
When a sequence is executed using Run Sequence, methods set in the Sample Log
Table are always loaded from the directory where the currently loaded method at the
Top level is saved.
>>
Run.
The Start Sequence dialog box appears.
Start Sequence Dialog Box
A data batch directory as the destination for the data files is created based on the
current date and time. The format is year, month (A-L corresponds to 1-12), day,
hour(A-X corresponds to 0-23), and a letter(00,01,02,..zy,zz). If you wish, you can
type your own directory name, instead. The name can be up to eight characters long.
>>
Load.