Agilent 7500 ICP-MS ChemStation Operator's Manual
Appendix
Select Use Dead Time Calibration.
4 Introduce the 50ppb Er solution.
166
Tune the sensitivity to make the count of the major isotope
Er in the range of
50,000 - 1,000,000 count / 0.1 sec.
A-31