Agilent 7500 ICP-MS ChemStation Operator's Manual
Startup, Shutdown and Status
The Ion Lenses Test dialog box will be displayed.
Ion Lenses Test dialog box
The ion lenses will be checked and the measured values displayed in the Inputs area
five seconds later.
Compare the set value in the Outputs area with the measured value in the Inputs
area. If the measured value is within +/-1 V of the set value, "OK" will be displayed
to the right of the measured value. Measured values out of this range will be marked
with "NG"
5 Click Close.
The Ion Lenses Test dialog box will be closed.

Nebulizer Test

This test checks whether the nebulizer pressure is in the appropriate range.
1 Select Instrument >> Instrument Control from the Top menu.
The ICP-MS Instrument Control window will be displayed.
2 Select Diagnostics >> Nebulizer Test.
3-20